It looks like you are proposing some kind of mixed approach, not a simple “less data, faster scan, ML to the rescue”. I understand how MRI works but you are surely and obviously much more knowledgeable than me, so I simply wish you luck!
My problem with the article comes from reading that someone telling about using generative models on health data, I don’t think is time for this yet.
Just a simple question: to achieve a guarantee that those results are at least equal or better than the ones we have now on our battle tested setups, shouldn’t we use the same sampling we use on a “default” MRI? I mean… using those reconstruction algorithms to try to achieve a better result, without downsampling so that a standard reconstruction can still be performed to be checked against?
It seems unlikely you wouldn’t appreciate this already, but clinical MRI has not fully sampled in a long time. Between the old and the new - reduced phase resolution (image plane and slice plane) parallel imaging, compressed sense (or sensing), reduced frequency resolution with partial echo techniques, high reconstruction max trim with low acquisition matrix, the list is quite long.
The changes in resulting artefacts as acceleration techniques change (eg high compressed sense values) is a bit of a change to how people work. Very digital looking artefacts are just gross.
Thanks for your work! We need more speed.
Watching the mental gymnastics used to deal with multi band/simultaneous multi slice (or whatever vendors call it) with functional MRI was impressive to me.
Using ML to work out voxel results in such low spacial resolution scans has got to be scary.
Note, I am biased because I research MRI acquisition and reconstruction methods and I am rolling out trials of fast MRI methods (that use some ML in the reconstruction) to find out how robust the methods actually are in practice.