It's easy to do. TEM/STEM vs SEM vs STM. All completely different things. This is what happens when scientists name things :P
For those confused:
TEM/STEM: An electron beam is transmitted through your sample. Good for atomic-scale imaging.
SEM: An electron beam is scanned across your sample, but none are transmitted through. Good for topography/surface features (the interaction volume of the beam is too large for atomic resolution).
STM: No electron beam. Instead think of a vinyl record player, and physically scanning a very sample tip across the surface of your sample. Good for atomic-scale imaging of a surface.
A good STM image and a good STEM image can, at first glance, look quite similar (especially for a 2D material like graphene), but they're very different techniques.
So, yeah, my comment is not entirely accurate about the electron densities of the atoms. If you feel it is too misguided, I'll remove it.
This is what happens when you ask a theoretical chemist to explain an experiment. ;)