There's a difference between simple ventilator [1], and an anesthesia machine [2] that adds gas mixing, scavenging, etc.
ICU patients are anesthetized using IV sedation (a common regimen in the US is fentanyl/propofol), not inhalational anesthetics. Most vents only have simple inline filters to reduce contamination.
[1] A classic vent in the US is the Puritan Bennett 840. Here's its manual showing filters: https://www.medtronic.com/content/dam/covidien/library/us/en...
Seriously thinking about putting together a group of people to start producing these locally. Around campus we have a significant capacity in manufacturing precision mechanical devices.
I read this as "how to save a life with a vaccuum cleaner in extremis", and not "here's a good ventilator design". Is it a good idea? I dunno. I'll check on the vaccuum today just in case.
Also FWIW: the air path doesn't seem like the critical piece here. My (amateur) intuition tells me that the number of lives saved by amateur ventilators will be rather less than the number killed by amateur intubation devices.
Again: I'd be sort of OK with a ventilator manufactured by a HN reader. Under no circumstances do I want a software geek putting a tube in me.
I would consider the following a bare-minimum feature set for a COVID patient ventilator, any less and it would do more harm than good:
- cycle between an inspiratory and expiratory phase
- during the inspiratory phase, deliver an adjustable volume of gas (in the region of 6 ml per kg of patient's body weight) using as little pressure as possible, with an adjustable upper limit of pressure (in the region of 30 cmH2O)
- during the expiratory phase provide an adjustable pressure against exhalation (in the region of 0-30 cmH2O)
- allow blending of air and oxygen to deliver an adjustable inspired oxygen fraction
- allow the timing of the inspiratory and expiratory phases to be independently varied, thereby allowing the respiratory rate and the ratio of inspiration to expiration time to be controlled. Permit respiratory rates in the range of 8-60 breaths per minute
- measure and display the pressures and volumes within the respiratory system
- allow adjustable alarm-limits for pressures and volumes, and provide clear audible and visual alarms if these values are exceeded
For added patient safety and benefit, the following would be helpful
- measure inspired and expired oxygen and carbon dioxide content, and display both on a continuous waveform graph
- allow the patient to initiate the inspiratory phase by sensing patient inspiratory effort and providing pressure support for inhalation; ie sense when the patient inhales and deliver 10-15 cmH2O pressure for 0.5 seconds to augment inhalation
Ask programmers, they’ll get you up to speed.
This was the standard ventilator in Europe for decades. It is simple to design, does not even require electricity (mechanical solution driven by air from central compressed air system) so all parts can be sourced locally, and the design has a clinically proven track record.
Here is a video with schematic showing how it works:
https://youtube.com/watch?v=CrUfblki6Zs