Re:A woman’s brain stimulator stopped working after her home... by seehafer on Sunday May 06, 2018 @12:57PM (#17007095)
I used to design these devices (deep brain stimulators).

As part of the design process for every FDA-regulated product, you are obliged to perform a risk analysis to judge the severity and probability of risks to a patient using your product.

One of the risks was “Exposed to a strong current that could destroy/disable the onboard electronics”. The design mitigation was a fail safe shutdown.

Re:A woman’s brain stimulator stopped working after her home... by Waterluvian on Sunday May 06, 2018 @01:09PM (#17007166)
Sounds like it worked well:

"Fortunately, when he and his colleagues checked the woman’s implant, they found that it had not been damaged. A safety function designed by the manufacturer had taken its cue from the sudden current running through the house and caused the device to switch itself off, pre-empting any interference from the lightning. The programming was intact; the battery still had plenty of charge.

When they fired up the device, it worked perfectly, none the worse for the storm."

Re:A woman’s brain stimulator stopped working after her home... by rsynnott on Sunday May 06, 2018 @01:22PM (#17007247)
Seems like bad UX and/or user instruction, really; they should presumably have the device warn of this happening, and tell end users how to turn it back on. For a critical health device like this, having to visit the doctor to be told “have you tried turning it on and off again” doesn’t seem ideal. What if it had been a pacemaker or something?
Re:A woman’s brain stimulator stopped working after her home... by tscs37 on Sunday May 06, 2018 @02:09PM (#17007483)
It's very hard to design a device that can reliably tell that it has not been damaged in a way that hurts the user.

We have this with regular socket power too; if live and ground are shorted, connecting the device will shut off power to prevent the user from any harm due to a charged casing. There is no feedback other than "everything stopped working" because the GFI cannot tell if the situation is cleared without putting up power and potentially harming the user.

Think of the other scenario, the stimulator gets damaged, switches back on and kills the user. Pacemaker could do the exact same, if it's damaged it could kill the user very easily.

Re:A woman’s brain stimulator stopped working after her home... by UncleEntity on Sunday May 06, 2018 @02:55PM (#17007745)
> Think of the other scenario, the stimulator gets damaged, switches back on and kills the user. Pacemaker could do the exact same, if it's damaged it could kill the user very easily.

I was talking to a guy who's implanted defibrillator failed a while back.

It shorted out and started shocking him as fast as the capacitor(?) could recharge. It was shocking him, his nephew, the EMTs and ER staff until they finally got the thing shut off. His doctor was saying he's really, really lucky to be alive.

So, yeah, fail safe is good.

Re:A woman’s brain stimulator stopped working after her home... by Maxious on Sunday May 06, 2018 @06:24PM (#17008968)
"If an ICD hasn't been turned off and starts to fire, holding a magnet right over the device will temporarily disable it and keep it from delivering multiple shocks. Some hospice programs give magnets to families who are taking care of a hospice patient with an ICD at home."

http://articles.chicagotribune.com/2012-03-28/health/sc-heal...

Re:A woman’s brain stimulator stopped working after her home... by phkahler on Sunday May 06, 2018 @03:26PM (#17007952)
Once again the auto industry is ahead of someone. No time to explain the safety built in to an airbag system. It's mind numbing.
Re:A woman’s brain stimulator stopped working after her home... by baby on Sunday May 06, 2018 @06:45PM (#17009061)
That's because of the number of death per day in the auto industry. They have a lot of free guinea pigs to experiment with.
Re:A woman’s brain stimulator stopped working after her home... by DrScump on Monday May 07, 2018 @04:23AM (#17011155)
Well, assuming you don't have the defective Takata airbags.
Re:A woman’s brain stimulator stopped working after her home... by jakeogh on Monday May 07, 2018 @04:56AM (#17011280)
The ethical solution to human life critical systems is to make them 100% open. The user should have complete access, even to things we suspect are not necessary, like masks themselves. Our courts can handle this. Demanding it once you need it wont work.
Re:A woman’s brain stimulator stopped working after her home... by Retric on Sunday May 06, 2018 @01:28PM (#17007272)
First it was not a critical life saving device so shut down was a completely viable option. Second, ideally you don't want people to have any interaction with these devices. If they can restart it they can probably mess with it in other less useful ways.

Really you want something like this to be as simple as possible without harming the patient.

Re:A woman’s brain stimulator stopped working after her home... by seehafer on Sunday May 06, 2018 @02:17PM (#17007529)
Instructions for Use indicate that in such an event you should immediately go see your physician.

A pacer is different than a brain stimulator. And pacers generally try to restart themselves if they can. Brain stimulators, because of the sensitivity of where you’re stimulating (a few cm, literally, away from regions that control breathing, heart rate, etc), the least risky option is a shut down.

Re:A woman’s brain stimulator stopped working after her home... by bastawhiz on Sunday May 06, 2018 @01:33PM (#17007294)
> they should presumably have the device warn of this happening

How? Shock your brain? It's an implant.

> What if it had been a pacemaker or something?

The same question stands. How do you make the device provide reasonable feedback? Make your heart skip a few beats? How exactly is a pacemaker going to warn a user about an error condition?

Given the circumstances, and the rarity of the error case, it sounds like the device switching itself off is the best course of action.

Re:A woman’s brain stimulator stopped working after her home... by dvno42 on Sunday May 06, 2018 @05:19PM (#17008611)
I can speak to this a bit. Some pacers can be programmed to beep in certain situations to indicate things like, low battery. Alternatively you can set some models up so that when a magnet is applied, different programmable actions occur such as, changing the heart beat rythm to indicate what the expected remaining battery life is. There are also "at home" base stations which upload stats to the manufacturer who can call the patient (through the base station) when there are unexpected problems.
Re:A woman’s brain stimulator stopped working after her home... by MertsA on Monday May 07, 2018 @01:59AM (#17010661)
>changing the heart beat rythm to indicate what the expected remaining battery life is.

Wow, now that is an interesting user interface.

Re:A woman’s brain stimulator stopped working after her home... by rsynnott on Sunday May 06, 2018 @01:40PM (#17007327)
There’s an external charger, per the article; it could presumably feed back to that when charging.
Re:A woman’s brain stimulator stopped working after her home... by icebraining on Sunday May 06, 2018 @02:05PM (#17007462)
Connecting your brain implant to the charger after it unexpectedly shut off due to lightning seems like a good way to earn a Darwin award.
Re:A woman’s brain stimulator stopped working after her home... by bastawhiz on Sunday May 06, 2018 @01:50PM (#17007371)
If the implant is shut off for safety reasons related to electrical issues, the last thing I'd want interacting with it in any way is the charger. Hell, we can't even get Macbooks that don't fry USB devices.
Re:A woman’s brain stimulator stopped working after her home... by microcolonel on Sunday May 06, 2018 @02:06PM (#17007469)
> Hell, we can't even get Macbooks that don't fry USB devices.

Not sure why you'd assume the mainboard and peripheral circuit design on MacBooks would be particularly good, they seem to make the same mistakes for years on end.

Re:A woman’s brain stimulator stopped working after her home... by marcosdumay on Sunday May 06, 2018 @01:45PM (#17007344)
If it was obvious that the device stopped working, they made the right choice. The device can not tell with certainty that it wasn't damaged, so going into a specialist is the correct action.
Re:A woman’s brain stimulator stopped working after her home... by rsynnott on Sunday May 06, 2018 @02:03PM (#17007450)
Yep, that’s fair enough, I suppose. What I’d wonder about would be, would it always be obvious?