News for hackers. Stuff that matters.

wormyrocks

The machine still works if you yank the 4-pin connector from the cellular board, so that's the easiest way. We gotta add that to the instructions...
Hi, I worked on this project and have a couple comments to address the issues people are raising.

- There are established medical protocols to treat COVID-19 patients with BiPAP machines, including the addition of a viral filter to mitigate aerosolizing the virus. People are using these protocols now and we link to them from the site.

- There are two separate firmware hacks presented. The first one modifies ~20 bytes and provides UI access to BiPAP code left in the existing binary, which would allow the more common CPAP machines to fulfill the same limited function. The second is a PoC of a 'full ventilator' mode.

- The manufacturer's CPAP and BiPAP lines have identical mainboard designs and a near identical array of sensors, which provide realtime data including tidal volume calculations. This project exists as a PoC to show that it would be possible - simple, even - for the manufacturer to convert CPAPs to BiPAPs via an OTA update - or, with significantly more effort, to fully featured ventilators. It is likely that they are reluctant to acknowledge this is possible in fear of destroying the market for the BiPAP machines.

- If you are a SleepyHead/OSCAR developer, or have access to an AirCurve S10 and an ST programmer, I would love to talk.